HomeAbout UsProductsDiseasesTechnologyResourcesCountry Go

Specifications | Model 515B Oximeter

PRINCIPLE OF OPERATION
Red/Infrared absorption.

SpO2 (OXYGEN SATURATION)
Range: 0-100%
Accuracy: ±2% SpO2 (for 80-100% SpO2) (1 SD, or 68% of readings within claim) Unspecified for 0-79% SpO2
Resolution: 1%
Averaging: 8 seconds

PULSE RATE
Range: 30-250 bpm
Accuracy: ±2 bpm
Resolution: 1 bpm
Averaging: 8 seconds

SENSORS
Reusable Y-Sensor™ (can be sterilized and used with all patient populations) and reusable adult finger sensor.


Top of Page

General

ALERT
Automatic and adjustable limits for SpO2 and Pulse Rate
Audio: Adjustable volume, 2 min. silence or OFF (LED indicators)
Visual: Flashing numerics upon violated limit(s) & red "Alert Bar". Indication of sensor disconnect, sensor off patient, low signal, insufficient light, light interference, pulse out of range, sensor faulty and monitor faulty conditions.

DISPLAY
Numerics: 7-Segment LED's
Pulse Bar: LED Pulse Bar Graph proportional to signal strength

PHYSICAL
Size: 3.3" H × 9.0" W × 8.0" D (8.38 × 22.86 × 20.32 cm)
Weight: 6 lb (2.72 kg)


Top of Page

 


Respironics Home  |  Glossary  |   Careers  |  Online Stores  |  Warranties  |  Providers
Suppliers  |  Clinicians  |  Media Resource Library  |  Contact Us
All Rights Reserved  |  Privacy Notice  |  Terms & Conditions  |  Site Map

 